The instrument should be calibrated before use. It should also be calibrated if there is a change in conditions (temperature, humidity, etc.) Generally, the calibration should be checked by taking a reading on the supplied standard at least every half-hour during continuous use.
Table of Contents
Related FAQs
FAQ Categories
How often should I calibrate the instrument?
Related products
Aesthetixยฎ: The Ultimate Surface Finish Solutionโ
- Fully characterise all aspects of appearance perception - Why does this matter?
- Modular solution: Surface Brilliance | Textureย | Effect Finish
- Aesthetixยฎ can be used in-laboratory, portable in-field, inline in the production environment or as part of a Cobot / Robot solution
- Measures flat surfaces, curved parts and small surface areas
- Affordable technology starts at $7,500
AFNOR Viscosity Flow Cup
- 2mm - 8mm orifice diameter
- 15 โ 3000 cSt viscosity range
- 25 -150 sec flow times
- Supplied with a certificate of conformity
- Sheen 419/2.5, 419/4, 419/6 & 419/8 equivalent
ASTM D1200 Viscosity Flow Cup
- 2.1mm - 5.8mm orifice diameter
- 10 โ 1200 cSt viscosity range
- 30-100 sec flow times
- Supplied with a certificate of conformity
- Sheen 406/1, 406/2, 406/3, 406/4 & 406/5 equivalent
Bespoke cutter die – 1 or 2 of the same shape
Bespoke cutter die - 1 or 2 of the same shape