If the instrument is used in the wrong range, it can become oversaturated. This occurs when measurements on mirror-like surfaces (130+GU @60ยฐ) are taken using the black range. To get accurate results on a mirror-like surface, the range must be set mirror or auto in the measurement menu.
My instrument always reads the same